SUPPORT - MANUALS & DOWNLOADS
Software that capture measurement data

Alwan Software (APV, APS)

Alwan Print Verifier X

Manual and Auto upload modes.

  1. Download ChromaChecker APV Job Template that corresponding your control strip, test form or Target (Download Section of the manual)— unzip the archive file.
  2. Launch APV application.
  3. Configure FTP connection if automatic upload is expected (not mandatory but recommended). To make FTP  connection  available — activate FTP account on the website:





    and enter exactly same values in APV Export Preferences:


    Press Open URL to check connection.

  4. Import required templates into the application – remember the application has to be in Admin mode - to enter this mode click on "Lock" icon. In Admin mode top application menu will have more options: 



     
  5. Go to Job Template section and by clicking import icon (on the screenshot below marked with green) add all ChromaChecker Job templates that reflect your Instrument/ Target combination. You may import more templates.



  6. You may want to edit some additional information like Print Company Name, Site, Operator name. 

  7. If you want to automatically upload data to ChromacChecker site you have to specify Print Device Name and Substrate (Print Device section of Job template). To edit press Ikon with a pencil. 



    In next step, depending on ChromaChecker tool you have to enter corresponding values in a specific way:





    Case 1 —  Instrument Inspector

    Go to Instrument Tab in Instrument Inspector.

    Click on preferences icon

     


    Enter corresponding values in Instrument/Target Preferences:





    Case 1 — Print Inspector

    Go to Track Preferences and setup Auto-recognition parameters.




  8. Perform measurements. Be sure that you connected Instrument and calibrated it before. 




  9. Upload files manually if automatic FTP method is off. Locate where files are stored and fdrag and  drop them on upload icon in a specyfic  ChromaChecker tool.

 

General scanning rules for spectral measurements

  • Proper measurements require temperature of 23 +/- 1°C (73 +/- 2°F), 40-60 % RH
  • Take some additional measurement before for internal warmup
  • For multiple measurements, always start by performing a new calibration.
  • Check if the instrument is scratching target - if so clean the bottom of the instrument and if necessary use paper bar to lift ruler slightly up - be careful not to lift it too much.
  • Clean calibration plate (ceramic tile) before calibration. Be sure that instrument is not very cold or hot (store it in its recommended temperature). Check if an optical patch is free of artifacts and dust.

 

Tips


If the fast upload is required ChromaChecker DataCollector software can be used instead of FTP connection. Simple select proper Input Folder. We advise using auto-recognition selection for Tracks. 

Clicking on the measurement on the list with right mouse button you can easy find measurement file location on the hard drive. Measurement file may be uploaded using drag and drop on upload icon in ChromaChecker website. More than one file may be uploaded in one step.